Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
933
User Manual 1178.8520.02 ─ 01
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<Offset>
<numeric value>, specified in Hz, kHz, MHz or GHz, rounded to
the nearest Hz
*RST:
0 Hz
Default unit: HZ
Example:
//Define an offset between generator output frequency and ana-
lyzer frequency
SOUR:EXT:FREQ:OFFS 10HZ
Manual operation:
See
"(Automatic) Source Frequency (Numerator/Denominator/
SOURce<si>:EXTernal<gen>:POWer[:LEVel]
<Level>
This command sets the output power of the selected generator.
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<Level>
<numeric value>
*RST:
-20 dBm
Default unit: DBM
Example:
//Define generator output level
SOUR:EXT:POW -30dBm
Manual operation:
See
SOURce<si>:EXTernal<gen>[:STATe]
<State>
This command activates or deactivates the connected external generator.
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation:
See
Setting Basic Measurement Parameters