Configuration
R&S
®
ESW-K58
32
User Manual 1179.0880.02 ─ 03
Note:
The "External Trigger 1" softkey automatically selects the trigger signal from the
"TRIGGER 1 INPUT" connector on the front panel.
For details, see the "Instrument Tour" chapter in the R&S
ESW Getting Started manual.
"External Trigger 1"
Trigger signal from the "TRIGGER 1 INPUT" connector.
"External Trigger 2"
Trigger signal from the "TRIGGER 2 INPUT / OUTPUT" connector.
"External Trigger 3"
Trigger signal from the "TRIGGER 3 INPUT / OUTPUT" connector on
the rear panel.
Remote command:
Trigger Level
Defines the trigger level for the specified trigger source.
For details on supported trigger levels, see the instrument data sheet.
Remote command:
TRIGger<tp>[:SEQuence]:LEVel[:EXTernal]
Trigger Offset
Defines the time offset between the trigger event and the start of the measurement.
Offset > 0:
Start of the measurement is delayed
Offset < 0:
Measurement starts earlier (pretrigger)
Remote command:
TRIGger<tp>[:SEQuence]:HOLDoff[:TIME]
Drop-Out Time
Defines the time that the input signal must stay below the trigger level before triggering
again.
Remote command:
Slope
For all trigger sources except time, you can define whether triggering occurs when the
signal rises to the trigger level or falls down to it.
Remote command:
Trigger Holdoff
Defines the minimum time (in seconds) that must pass between two trigger events.
Trigger events that occur during the holdoff time are ignored.
Remote command:
TRIGger[:SEQuence]:IFPower:HOLDoff
Trigger