Remote commands for Multi CISPR APD measurements
R&S
®
ESW-K58
86
User Manual 1179.0880.02 ─ 03
Parameters:
<Value>
numeric value
Default unit: dB
Example:
SENS:MAPD:CHAN1:EMAR 1.8
Manual operation:
See
[SENSe:]MAPD:CHANnel<n>:EVALuation
<Value>
This command enables the evaluation table.
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
Example:
SENS:MAPD:CHAN1EVAL ON
Manual operation:
See
[SENSe:]MAPD:CHANnel<n>:INSert
<Direction>
Inserts a channel into the evaluation table before or after the selected channel.
Suffix:
<n>
.
1..n
Channel
Setting parameters:
<Direction>
BEFore | AFTer
Example:
SENS:MAPD:CHAN1:INS BEFore
Usage:
Setting only
Manual operation:
See
[SENSe:]MAPD:CHANnel<n>:LOAD
<FileName>
Loads an evaluation table configuration. Evaluation tables are stored in an xml file. The
default directory for evaluation lists is C:\R_S\INSTR\user\MAPD.
Suffix:
<n>
.
1..n
irrelevant
Parameters:
<FileName>
String containing the file name.
Example:
SENS:MAPD:CHAN1:LOAD 'test.xml'
Manual operation:
See
Configuration