Remote commands for Multi CISPR APD measurements
R&S
®
ESW-K58
89
User Manual 1179.0880.02 ─ 03
For a single measurement, the R&S
ESW stops measuring when it has reached the
end frequency. When you start a continuous measurement, it stops only if you abort it
deliberately.
If you are using trace modes MAXHold, MINHold and AVERage, previous results are
reset when you restart the measurement.
●
Single measurements
Synchronization to the end of the measurement is possible with
*OPC
,
*OPC?
or
*WAI
.
●
Continuous measurements
Synchronization to the end of the measurement is not possible.
It is thus recommended to use a single measurement for remote controlled mea-
surements, because results like trace data or markers are only valid after synchro-
nization.
Suffix:
<mt>
.
INITiate1 initiates a bargraph measurement.
INITiate2 initiates a scan.
Example:
//Start a single scan (with a scan count = 20), and wait until the
measurement is done
INIT2:CONT OFF
SWE:COUN 20
INIT2;*WAI
Usage:
Event
Manual operation:
See
See
8.6
Analysis
The functionality to analyze measurement results is the same as that of the Spectrum
application.
For a comprehensive list and description of remote commands, refer to the corre-
sponding topics in the user manual of the R&S
ESW.
8.6.1
Trace export
............................................................................................... 90
................................................................................................90
.................................................................................................90
.......................................................................................... 91
...........................................................................................91
Analysis