Welcome to the Multi CISPR APD application
R&S
®
ESW-K58
11
User Manual 1179.0880.02 ─ 03
1
= Channel bar for firmware and measurement settings
2+3 = Window title bar with diagram-specific (trace) information
4
= Diagram area
5
= Diagram footer with diagram-specific information, depending on result display
6
= Instrument status bar with error messages and date/time display
Channel bar information
In the Multi CISPR APD application, the R&S
ESW shows the following settings:
Table 2-1: Information displayed in the channel bar in the application for Multi CISPR APD measure-
ments
Ref Level
Reference level
Att
RF attenuation applied to input
Input
Input toggle state
ABW
Analysis bandwidth
AQT
Measurement time for data acquisition.
PS
Current state of the preselector.
Tuned Frequency
Frequency of the current measurement results
Limits
PASS or FAIL of defined limits.
SGL
Indicates the progress of single measurements.
The first number is the current measurement. The second number is the
total number of measurements.
Only displayed for single measurements and if the scan count is greater
than 1.
Frequency
Receiver Frequency
Understanding the display information