Remote Control Commands
R&S
®
NRP18S-xx
76
User Manual 1178.3686.02 ─ 05
Parameters:
<time>
Range:
0.00 to 0.10
*RST:
0.00
Default unit: Seconds
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID[:STATe]
<state>
Enables or disables the blanking out of time intervals in the timeslots.
Parameters:
<state>
*RST:
OFF
8.7.4
Trace Measurement
The trace measurement is used to acquire the course of power over a certain time.
During the measurement time (
measurements are made and the result is returned to the user as an array of values
with a predefined size
. The length of an individ-
ual measurement(-point) is determined from the ratio of measurement time and mea-
surement points. The entire result is called a "trace". Each trace must be triggered sep-
arately.
Remote commands:
[SENSe<Sensor>:]TRACe:AVERage:COUNt
...................................................................... 76
[SENSe<Sensor>:]TRACe:AVERage:TCONtrol
...................................................................76
[SENSe<Sensor>:]TRACe:AVERage[:STATe]
..................................................................... 77
..................................................................................... 77
[SENSe<Sensor>:]TRACe:MPWidth?
................................................................................ 80
[SENSe<Sensor>:]TRACe:OFFSet:TIME
........................................................................... 80
.....................................................................................80
[SENSe<Sensor>:]TRACe:REALtime
.................................................................................80
........................................................................................81
[SENSe<Sensor>:]TRACe:AVERage:COUNt
<count>
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
Averaging is only effective, if
[SENSe<Sensor>:]TRACe:AVERage[:STATe]
ON
is
set.
Parameters:
<count>
Range:
1 to 65536
*RST:
4
[SENSe<Sensor>:]TRACe:AVERage:TCONtrol
<mode>
Available in trace mode.
Defines how the measurement results are output. This is called termination control.
Configuring the Measurement Modes