![R&S 1424.6721.02 User Manual Download Page 56](http://html1.mh-extra.com/html/rands/1424-6721-02/1424-6721-02_user-manual_774880056.webp)
Remote Control Commands
R&S
®
NRP18S-xx
56
User Manual 1178.3686.02 ─ 05
Example: Repeating termination control
Further settings for this example:
●
[SENSe<Sensor>:]AVERage:TCONtrol
REPeat
Every chopper phase is started by a trigger event and lasts the defined aperture time.
During a chopper phase, the trigger synchronization is high (
ON
). After 2 chopper phases, 1 measurement is completed. As defined by the average
count, after 2 measurements, the result is averaged and available.
1
7
3
4
5
6
2
1 = Start of the measurement cycle
2 = Trigger event
3 = Noninverted chopper phase
4 = Inverted chopper phase
5 = Measurement result
6 = Trigger synchronization
7 = Return to the start of the measurement cycle
Controlling the Measurement