![R&S 1424.6721.02 User Manual Download Page 152](http://html1.mh-extra.com/html/rands/1424-6721-02/1424-6721-02_user-manual_774880152.webp)
Remote Control Basics
R&S
®
NRP18S-xx
152
User Manual 1178.3686.02 ─ 05
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
Upper limit fail
0
0
0
0
0
0
0
0
0
0
0
0
0
0
+
Figure 10-14: Operation upper limit fail status registers
Querying the register:
●
STATus:OPERation:ULFail:CONDition?
●
STATus:OPERation:ULFail[:SUMMary][:EVENt]?
Table 10-16: Used operation upper limit fail status bits and their meaning
Bit no.
Short description
Bit is set if
1
Upper limit fail
Measured value exceeds the upper limit value.
Status Reporting System