![R&S 1424.6721.02 User Manual Download Page 49](http://html1.mh-extra.com/html/rands/1424-6721-02/1424-6721-02_user-manual_774880049.webp)
Remote Control Commands
R&S
®
NRP18S-xx
49
User Manual 1178.3686.02 ─ 05
Chapter 8.7.1, "Continuous Average Measurement"
●
Burst average (
"POWer:BURSt:AVG"
): The integration time of a measurement is
not predefined but determined by the sensor with the aid of a burst detector. The
start of a burst is detected when the measurement signal rises above a set trigger
level. The measurement ends when the signal drops below a trigger threshold.
See also
Chapter 8.7.2, "Burst Average Measurement"
●
Timeslot average (
"POWer:TSLot:AVG"
): The power is simultaneously measured
in up to 32 time slots. The measurement result is represented by a vector that can
contain up to 32 indices and contains the power of a time slot at each index.
See also
Chapter 8.7.3, "Timeslot Average Measurement"
●
Trace (
"XTIMe:POWer"
): A sequence of measurements is performed.
Chapter 8.7.4, "Trace Measurement"
8.5
Controlling the Measurement
The power sensor offers a bunch of possibilities to control the measurement:
●
Do you want to start the measurement immediately after the initiate command or
do you want to wait for a trigger event?
●
Do you want to start a single measurement cycle or a sequence of measurement
cycles?
●
Do you want to output each new average value as a measurement result or do you
want to bundle more measured values into one result?
8.5.1
Starting and Ending a Measurement
ABORt
Immediately interrupts the current measurement. If the measurement has been started
as a single measurement (
), the sensor goes into the idle
state. However, if a continuous measurement is in progress (
ON
), the trigger system of the sensor enters the waiting for trigger state, and if the trig-
ger condition is met, a new measurement is immediately started.
See also
Chapter 8.5, "Controlling the Measurement"
Usage:
Event
Controlling the Measurement