Remote Control Commands
R&S
®
NRP18S-xx
68
User Manual 1178.3686.02 ─ 05
SENS:FUNC
Possible CALC:FEED
Meaning
"POWer:RANDom"
Randomly selected value from the
measurement interval
"XTIMe:POWer"
"POWer:TRACe"
Measurement sequence
"POWer:PEAK:TRACe"
Peak value of the samples per
trace point
"POWer:RANDom:TRACe"
Randomly selected value of the
samples per trace point
Parameters:
<mode>
*RST:
"POWer:AVERage"
Example:
The following sequence of commands configures a peak trace
measurement:
*RST
SENSe:FUNCtion "XTIMe:POWer"
SENSe:FREQuency 1.0e9
SENSe:TRACe:POINts 500
SENS:TRAC:TIME 20e-3
TRIGger:SOURce INTernal
TRIGger:SLOPe POSitive
TRIGger:DTIMe 0.001
TRIGger:HYSTeresis 0.1
TRIGger:LEVel 30e-6
SENSe:TRACe:AVERage:COUNt 8
SENSe:TRACe:AVERage:STATe ON
CALCulate:FEED "POWer:PEAK:TRACe"
INITiate
FETCh?
[SENSe<Sensor>:]AUXiliary
<mode>
Enables the measurement of additional measured values that are determined together
with the main measured value.
Parameters:
<mode>
NONE | MINMax | RNDMax
NONE
No additional values are measured.
MINMax
Minima and maxima of the trace are transmitted together with
the measured value.
Usually, extreme values are lost due to averaging the measured
values.
RNDMax
Randomly selected samples are transmitted. All evaluations use
these values instead of the average values.
*RST:
NONE
Configuring and Retrieving Results