[SENSe<Sensor>:]TIMing:EXCLude:STARt
and
, see
Chapter 8.8.3, "Setting Exclusion Time"
Remote commands:
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:COUNt
................................................................ 75
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:WIDTh
.................................................................75
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID:OFFSet[:TIME]
..............................75
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID:TIME
............................................75
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID[:STATe]
........................................ 76
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:COUNt
<count>
Sets the number of simultaneously measured timeslots in the timeslot mode (see
Parameters:
<count>
Range:
1 to 128
*RST:
8
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:WIDTh
<width>
Sets the length of the timeslot (see
Parameters:
<width>
Range:
10.0e-6 to 0.10
*RST:
1.000e-3
Default unit: Seconds
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID:OFFSet[:TIME]
<time>
Determines the distance from the start of the timeslots to the start of the interval to be
blanked out (see
Parameters:
<time>
Range:
0.00 to 0.10
*RST:
0.00
Default unit: Seconds
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID:TIME
<time>
Sets the length of the time interval in the timeslots to be excluded from the measure-
ment (see
). The parameter applies to each individual timeslot.
Note:
Even if the exclusion interval exceeds the timeslot because, for example, its right
limit is outside the timeslot, correct results are obtained. In the extreme case, where
the interval length has been set to a value greater than the timeslot length, 0 W is out-
put as the measured power. No error message is output.
Configuring the Measurement Modes