Remote Control Commands
R&S
®
NRP18S-xx
85
User Manual 1178.3686.02 ─ 05
8.8.2
Setting the Frequency
The frequency of the signal to be measured is not automatically determined. For ach-
ieving better accuracy, the carrier frequency of the applied signal must be set.
[SENSe<Sensor>:]FREQuency
<frequency>
Transfers the carrier frequency of the RF signal to be measured. This frequency is
used for the frequency-response correction of the measurement result.
The center frequency is set for broadband signals, e.g. spread-spectrum signals, multi-
carrier signals.
Parameters:
<frequency>
Range:
0.0 to 110.0e9
*RST:
50.0e6
Default unit: Frequency
Example:
FREQ 10000
8.8.3
Setting Exclusion Time
In the burst average and the timeslot average modes, you can define a time period at
the beginning or at the end of an integration period, which is excluded from the mea-
surement result, as shown in
Figure 8-6: Effect of commands SENS:TIM:EXCL:STAR and SENS:TIM:EXCL:STOP
Remote commands:
[SENSe<Sensor>:]TIMing:EXCLude:STARt
........................................................................ 86
[SENSe<Sensor>:]TIMing:EXCLude:STOP
.........................................................................86