Remote Control Commands
R&S
®
NRP18S-xx
67
User Manual 1178.3686.02 ─ 05
FETCh<Sensor>:ARRay[:POWer][:AVG]?
Queries the last valid measurement result of a measurement with enabled data buffer
mode.
Usage:
Query only
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
Queries the last valid measurement result.
Usage:
Query only
FETCh<Sensor>[:SCALar][:POWer]:BURSt?
Queries the last valid measurement result for the burst average measurement mode.
Usage:
Query only
FETCh<Sensor>[:SCALar][:POWer]:TSLot?
Queries the last valid measurement result of the timeslot average measurement mode.
Usage:
Query only
CALCulate:FEED
<mode>
If you query measurement data using
FETCh<Sensor>[:SCALar][:POWer][:
, the power sensor returns data of the measurand that was configured before.
Generally, this measurand is the average power. However, the power sensor can also
output data of other measurands.
To configure which measurand the
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
command sends, use the
CALCulate:FEED
command before the measurement is ini-
tiated. Depending on the measurement mode, the following settings are possible:
SENS:FUNC
Possible CALC:FEED
Meaning
"POWer:AVG"
"POWer:AVERage"
Average value
"POWer:PEAK"
Peak value
"POWer:RANDom"
Randomly selected value from the
measurement interval
"POWer:BURSt:AVG"
"POWer:AVERage"
Average value
"POWer:PEAK"
Peak value
"POWer:RANDom"
Randomly selected value from the
measurement interval
"POWer:TSLot:AVG"
"POWer:AVERage"
Average value
"POWer:PEAK"
Peak value
Configuring and Retrieving Results