[SENSe<Sensor>:]TRACe:MPWidth?
Queries the attainable time resolution for the Trace mode. The result is the smallest
possible distance between two pixels, i.e. it is the smallest time interval that can be
assigned to a pixel.
Usage:
Query only
[SENSe<Sensor>:]TRACe:OFFSet:TIME
<time>
Adds an offset to the beginning of the trace sequence. Thus, the trace in the result dis-
play is moved in positive or negative x-direction. If you measure with more than one
sensor, you can use this offset to arrange the traces to each other. The start of record-
ing relative to the trigger event is set using
.
Parameters:
<time>
Range:
- (500 ms + TRGger:DELay) to 100 s
*RST:
0.0
Default unit: Seconds
Example:
TRAC:OFFS:TIME 1.0
[SENSe<Sensor>:]TRACe:POINts
<points>
Sets the number of required values per trace sequence.
Parameters:
<points>
Range:
1 to 100000
*RST:
260
[SENSe<Sensor>:]TRACe:REALtime
<state>
Available in trace mode.
If disabled, each measurement from the power sensor is averaged. If enabled, only
one sampling sequence per measurement is recorded, thus increasing the measure-
ment speed. With a higher measurement speed, the measured values of an individual
measurement are immediately delivered.
The averaging filter is not used, so the following settings are ignored:
●
[SENSe<Sensor>:]TRACe:AVERage[:STATe]
●
[SENSe<Sensor>:]TRACe:AVERage:COUNt
Parameters:
<state>
*RST:
OFF