PROGRAMMING THE AREARAE
4-41
5. “1.00” is the calculated correction factor of the selected gas
in Step 4.
TOX1
VOC
TOX2
Methane
LEL Factor = 1.00?
LEL
OXY
TOX1
VOC
Gamma
Methane
LEL Factor = 1.00?
LEL
OXY
AreaRAE
AreaRAE Gamma
6. If you do not want to modify the LEL correction factor,
press [Y/+] and go to Step 8. To modify this factor, press
[N/-] first. Then starting from the left-most digit, use the
[Y/+] or [N/-] to change the digit value and press [MODE]
momentarily to advance to the next digit. The flashing digit
will move on to next digit to its right. Repeat this process
until all 4 digits of the new factor are entered. Press and
hold [MODE] for 1 second to exit data entry mode. If there
is any change to the existing value, the display shows
“Save?” Press [Y/+] to accept the new value and exit the
gas selection submenu. Press the [N/-] to discard the
changes.
7. If the VOC sensor is installed, the display shows:
TOX1
VOC
TOX2
VOC Gas = ?
Isobutylene
LEL
OXY
TOX1
VOC
Gamma
VOC Gas = ?
Isobutylene
LEL
OXY
AreaRAE
AreaRAE Gamma
Otherwise, the message “No VOC installed” is displayed
and moves to the next submenu.
8. If you do not want to change the VOC measurement gas,
press [Y/+] to accept the current gas and go to Step 12.
9. If you want to modify the VOC measurement gas, press
[N/-] first, then use [Y/+] or [N/-] to scroll through a list of
gas names until the desired gas name appears in the LCD
display, then press [MODE] to select the new gas name.
10. The display shows “Save?” To confirm the new gas, press
[Y/+] to accept the change. Press [N/-] or [MODE] to
discard the change and move to the next step.