OPERATION OF THE AREARAE
2-26
Other Datalog options automatically start and stop:
Datalogging Event
Each time a datalogging operation is initiated, a datalog event
is created. Information, such as start time, datalogging period,
alarm limits, etc., is recorded in the event header, followed by
the measurement data.
Datalogging Pause
Datalogging automatically pauses under the following
conditions:
1.
Upon entering Program Mode.
Datalogging resumes
upon exiting the Program Mode.
2.
Upon entering PC Communication standby mode.
Datalogging resumes upon exiting PC communication
standby mode, if datalogging mode was not set to manual
start/stop.
In both cases a new datalog event will be created when the
datalogging is resumed.