OPERATION OF THE AREARAE
2-9
2.3 Turn Power On/Off
To turn ON
Press [MODE]. The audio buzzer beeps once, and the screen
displays the following, in order:
•
“ON!”
•
“Multi-gas Monitor Version n.nn” or “Multi-gas Radiation
Monitor Version n.nn” (the software version)
•
Custom power-on-name, which can be changed by using
the ProRAE Suite Software
•
Model number, serial number, current date and time, and
temperature
•
Each sensor socket to check if a valid sensor is installed.
If a new sensor is installed, a message to remind the user
to calibrate the sensor will be displayed followed by a
message that the alarm limits have been reset to the
default values for the new sensor.
•
Warranty expiration date for each sensor
•
Preset alarm limits for each sensor
•
Last calibration date
•
Battery
voltage
•
Shut off voltage
•
User
mode
•
Alarm
mode
•
Available data storage memory (in hours)
•
Datalog
mode
•
Datalog
period
•
Fresh air calibration if Power On Zero is enabled
•
Instantaneous reading of the gas concentration in ppm
To turn OFF
Press and hold [MODE] for 5 seconds. The monitor beeps
once every second during the power-down sequence. A
countdown timer shows the remaining seconds. After that, the
screen flashes “Off!” and then goes blank, indicating the
monitor is off.