PROGRAMMING THE AREARAE
4-22
4.6.2 Change Datalog Period
The datalog period can be programmed from 1 to 3,600
seconds (1 hour).
1. “Change Datalog Period?” is the third item in the Datalog
Program submenu item.
2. Press [Y/+], and the display shows “New Period = 0060”
with the left-most digit flashing, where “0060” is the
previously stored data log period.
3. To modify this period, start from the left-most digit and use
the [Y/+] or [N/-] to change the digit value. Then press
[MODE] momentarily to advance to the next digit. The
flashing digit advances to the next digit to the right. Repeat
this process until all four digits of the new period are
entered. Press and hold [MODE] for 1 second to exit data
entry mode. If there is any change to the existing value, the
display shows “Save?” Press [Y/+] to accept the new value
and exit the datalog submenu. Press [N/-] to discard the
changes and move to the next menu option.
4. To preserve the previously stored period, press and hold
[MODE] for 1 second. The monitor exits data entry mode
and moves to the next submenu item.