- 33 -
-
On placing the determined weight on the weighing pan accept it by pressing
Enter
key.
-
On completing adjustment process the balance returns to displaying submenu P1 -
Adjustment.
-
Return to weighing mode – in accordance with point 7.1.2.
It is recommended that the mass of an external adjustment weight is
approximately ¾ of balance’s maximum capacity.
If the DRH function is enabled in balance settings, then external adjustment
process is disabled.
9.4. Adjustment report printout
On completion of any type of adjustment process, the balance enables preparing a report
from adjustment process. The report can be printed on a connected printer and sent to a
computer and saved in a form of file for records.
P1 08 Report printout :
1: yes – report printout enabled
P1 08 Report printout :
0: no
– report printout disabled
Remember, that if the parameter is set for 1 (
YES)
, then a report is generated and sent
automatically.
Fig. 17. Submenu: Adjustment
The content of the report from adjustment process depends on settings of GLP
parameters. Any option in the GLP submenu which attribute is YES is included in a report
from adjustment process.
Summary of Contents for PS 1000/X