- 84 -
Function
SWITCH ON THE BALANCE
Command
O1 CR LF (equal to pressing ON/OFF button)
Function
DISABLE ZUTOZERO
Command
A0 CR LF
Function
ENABLE AUTOZERO
Command
A1 CR LF
If a non-existing or incorrect command finished with CR LF is sent to a balance, it
responses with E S CR LF. Spaces between characters should be omitted, as they are
added only for the purpose of proper legibility.
26. ERROR MESSAGES
Message
Error
code
Error description
"Control sum error."
1.1
Error related to data transmission
" A/D converter error"
1.2
Converter error
"Measuring range
exceeded"
2.1
Maximum capacity (measuring range) of the
balance exceeded (over load)
"Measuring range
exceeded"
2.2
Maximum capacity (measuring range) of the
balance exceeded (over load)
"A/D Null"
2.3
No divisions from the converter
"A/D Full"
2.4
Maximum quantity of divisions from
converter exceeded
"Tare/Zero out of range"
2.5
Permissible value of zeroing or tarring range
exceeded
"Tare out of range"
2.6
Permissible value of tarring range exceeded
"Zero out of range”
2.7
Permissible value of zeroing range
exceeded
"Result > 4% Max"
2.8
Too high start mass (balance started with a
load on the weighing pan)
"Result > 1% Max"
2.9
The difference between mass of a currently
measured adjustment weight and mass
saved in balance memory is greater than
(difference >1%)
"Part < 1 Div"
2.10
Mass value of a single part in Parts Counting
mode is smaller than balance reading unit
“Part < 10 Div"
2.11
Load value placed on the weighing pan
while determining mass a single part in Parts
Counting mode is smaller than 10 reading
units of a balance
"Ref < 1000 Div"
2.12
Value of a reference mass in Percent Setup
mode in smaller than 1000 reading units
"Out of range"
3.1
Parameter value exceeds permissible value
"Incorrect value "
3.2
Impermissible parameter value
"DRH - locked"
3.3
Parameter change blocked (
DRH
function in
the factory menu is enabled)
Summary of Contents for PS 1000/X
Page 2: ...2 APRILE 2013...