- 38 -
12. SETTING BALANCE OPERATING PARAMETERS
Balance PS/X series, in menu group <P4 Readout> enables adjusting balance operation
to current ambient conditions at a workstation (filter) and required user needs (display
refreshment, autozero, displaying last digit).
Fig. 24. Submenu Readout – internal settings
12.1. lter settings
Depending on the ambient conditions at a workstation, the balance enables setting the
filtering value. In case of very good operating conditions it is recommended to set the
filtering value to very fast (parameter value 01 Filter set to 1). If the operating conditions
are harsh (air drafts, vibrations) set the filter to slow or very slow (parameter value 01 Filter
set to 4 or 5). The effectiveness of filter operation differs in relation to the measuring range.
The filter operates with lower accuracy while the mass indication is quickly increasing after
placing a load on the weighing pan. Filter accuracy is increased when weighed mass is
within filter’s set operation range (parameter: filter operation range is available only in
balance’s service menu, and it is inaccessible for the user). Depending on filter settings the
weighing time is either shorter (low filter values) or longer (high filter settings).
12.2. Value release
Select one of available value release options: fast, fast+reliable or reliable. Depending on
accepted criterion the weighing time will be shorter or longer.
12.3. Time interval of display refreshment
The parameter determines time interval in which display indication is refreshed.
In case of the higher refreshment values, the display does not indicate intermediate and
unstable mass values occurring while loading and unloading weighed mass on the
weighing pan. For low refreshment values the display indicates any changes in the value
of weighed mass – which is required while dosing loose or liquid materials. The time
interval of display refreshment is set in seconds.
Summary of Contents for PS 1000/X
Page 2: ...2 APRILE 2013...