- 47 -
The display indicates mass of counted parts, that are loaded on balance’s weighing pan
(i.e. 10 parts). If the added amount of parts is below the currently counted one, then the
software automatically corrects mass of a single part. In this case it is APW = 5.2282
corrected to 5.1837. from now on the following parts are counted according to the new
mass of a single part.
This means enables counting mass of a single part from a standard quantity.
The software comprises four conditions for operation of
Automatic Accuracy Correction function
1. After adding the number of parts placed on balance’s weighing pan must be
greater than before adding
2. After adding the number of parts loaded on balance’s weighing pan must be less
than twice the quantity which was indicated on the display before adding
3. current quantity of parts must contain within a tolerance
±
0,3 of the total value,
4. stable measurement result.
If a user claims, that standard quantity is sufficient, it is possible to save the mass of a
single part in balance’s memory by pressing
RIGHT ARROW KEY
on balance’s overlay.
Fig. 33. Automatic Accuracy Correction – saving single part mass in the
database
Set the cursor next to a desired field and insert name for weighed parts. Press
Enter
key
(for saving the name)
and
Enter
key (for saving the value
)
. Next to the inserted name there
is mass of a single part. Now the record in the database is ready for recalling by using field
02 Recall standard.
Summary of Contents for PS 1000/X
Page 2: ...2 APRILE 2013...