1256 User Manual Publication No. 980855 Rev. A
Advanced SCPI Operations 8-16
EADS North America Test and Seervices
and end with a number between 0 and 100. The following
are valid state names:
STATE0
STATE7
State53
State100
For example, the commands:
PATH:DEF example,(@7(0,5,10,13))
SCAN
(@1(323),4(0:2),5(8:5),example,
1(0),state14,1(224))
Define a path name (“example”) and then define a SCAN list. Each
time a trigger is received, the next channel, path, or state in the
scan list is closed after the previous channel or path is opened.
Using the SCAN command example above, the following
sequence of actions occurs:
A) After a trigger is received, channel 323 of module 1 is closed
B) After the next trigger, channel 1(323) is opened, channel 4(0)
is closed
C) After the next trigger, channel 4(0) is opened, channel 4(1) is
closed
D) After the next trigger, channel 4(1) is opened, channel 4(2) is
closed
E) After the next trigger, channel 4(2) is opened, channel 5(8) is
closed
F) After the next trigger, channel 5(8) is opened, channel 5(7) is
closed
G) After the next trigger, channel 5(7) is opened, channel 5(6) is
closed
H) After the next trigger, channel 5(6) is opened, channel 5(5) is
closed
I) After the next trigger, channel 5(5) is opened, and all
channels defined for the path “example” are closed. (Module
7, channels 0, 5, 10, and 13)
J) After the next trigger, all the channels defined for the path
“example” are opened, and channel 1(0) is closed
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com