MPS 3 System – Troubleshooting
276
UI Internal Error
Call Service
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
277
Elevated internal UI temperature
Check for obstructed air flow
Maintenance required
CONFIRM
* Ensure the device has
sufficient air flow to cool
* Call service after multiple
occurrences
278
Internal UI temperature Error
Check for obstructed air flow
Unable to proceed. Call for Service
SHUTDOWN
* Restart and resume case
* Call service after multiple
occurrences
279
UI Voltage Critical
SHUTDOWN RESTART
* Restart and resume case
* Call service after multiple
occurrences
280, 281,
282, 283
Internal ECG Error
ECG cannot be displayed
CONFIRM
* Continue as normal. Internal
ECG is not essential to
operation.
* Call service after multiple
occurrences
284
Copy to Media Failed
CONFIRM
* Retry
* Try another Flash memory
device
285
User default Protocol Error
Resetting to Factory Defaults
CONFIRM
* Continue as normal. Check
Protocols in Protocol Manager
*Re-download Master File
* Call service after multiple
occurrences
286
Protocol error. Deleting all Protocols.
Copy or re-create the protocols if needed
CONFIRM
* Continue as normal. Check
Protocols in Protocol Manager
* Re-download Master File
* Call service after multiple
occurrences
Alarm Code
Message Display
Corrective Action
30
Summary of Contents for MPS 3
Page 1: ...Quest Medical Inc...
Page 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Page 12: ...MPS 3 System Introduction 6...
Page 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Page 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Page 60: ...MPS 3 System System Installation 18...
Page 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Page 90: ...MPS 3 System Startup This page intentionally left blank 9...
Page 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Page 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Page 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Page 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Page 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Page 178: ...MPS 3 System Functional Use 32...
Page 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Page 231: ...MPS 3 System Troubleshooting 40...
Page 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Page 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Page 234: ...MPS 3 System Troubleshooting 43...
Page 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Page 252: ...MPS 3 System Service 4...
Page 256: ...MPS 3 System Part Numbers and Accessories 4...
Page 258: ...MPS 3 System Part Numbers and Accessories 6...
Page 259: ...MPS 3 System Part Numbers and Accessories 7...
Page 260: ......
Page 268: ...MPS 3 System Appendix This page intentionally left blank 8...