MPS 3 System – Troubleshooting
185
Air detected in delivery line. Flow continues
Manually check for air
CONFIRM
* Check delivery line to ensure it
is fully inserted in sensor
* If sensor is ignored, manually
inspect delivery line for air
moving forward
186
System Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
187
Max System pressure
Check circuit
CONFIRM
* Check the lines for clamps,
kinks, or occlusions
* Manually actuate delivery and
vent valves several times.
* Check heat exchanger to
ensure it is secured tightly with
the locking knob
* Restart System
* Contact service after multiple
occurrences
188
Max external Antegrade pressure
Check circuit
CONFIRM
* Check the lines for clamps,
kinks, or occlusions
* Check the stopcock position at
the sensor
* Try to reestablish zero by re-
zeroing sensor
* If it is believed sensor is faulty,
replace sensor and try again.
* Contact service after multiple
occurrences
189
Max external Retrograde pressure
Check circuit
CONFIRM
* Check the lines for clamps,
kinks, or occlusions
* Check the stopcock position at
the sensor
* Try to reestablish zero by re-
zeroing sensor
* If it is believed sensor is faulty,
replace sensor and try again.
* Contact service after multiple
occurrences
190
Unable to maintain flow rate setting
Check circuit
CONFIRM
* Check blood/crystalloid source
lines and pressures
* Stop flow and restart flow
* If problem persists, restart
system
* Call service after multiple
occurrences
Alarm Code
Message Display
Corrective Action
24
Summary of Contents for MPS 3
Page 1: ...Quest Medical Inc...
Page 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Page 12: ...MPS 3 System Introduction 6...
Page 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Page 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Page 60: ...MPS 3 System System Installation 18...
Page 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Page 90: ...MPS 3 System Startup This page intentionally left blank 9...
Page 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Page 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Page 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Page 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Page 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Page 178: ...MPS 3 System Functional Use 32...
Page 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Page 231: ...MPS 3 System Troubleshooting 40...
Page 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Page 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Page 234: ...MPS 3 System Troubleshooting 43...
Page 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Page 252: ...MPS 3 System Service 4...
Page 256: ...MPS 3 System Part Numbers and Accessories 4...
Page 258: ...MPS 3 System Part Numbers and Accessories 6...
Page 259: ...MPS 3 System Part Numbers and Accessories 7...
Page 260: ......
Page 268: ...MPS 3 System Appendix This page intentionally left blank 8...