MPS 3 System – Alarm Overview
10. Alarm Overview
Alarms and errors are the notifications to the operator to ensure safe and proper function of the
MPS 3 System. The alarm message states the alarm or error code, a brief description of the alarm,
and prompts for clearing the alarm. All Alarms are recorded in the Alarm Report and may be
accessed from the Menu option. The Alarm Report is retained permanently during System power
down and during total loss of power events. The oldest alarms are discarded in the event of the
Alarm log memory becoming full.
The AUDIO PAUSE button
is displayed in the High and Medium Priority Alarm screens. When
the AUDIO PAUSE button is selected, the audible alarm tone will be silenced for 30 seconds.
10.1. High Priority Alarms
High priority alarms halt fluid delivery and require acknowledgement by the operator. High
priority alarms are announced in a red message window with a flashing yellow border and a
repeating audible tone. When a high priority alarm is displayed, the display outside the
message window is dimmed and the flow knob and all buttons outside the high priority
message window become inactive until CONFIRM button is pressed.
The operator has the option to pause the audible alarm tone for 30 seconds.
The operator has to acknowledge the alarm by selecting an option in the Alarm message
window.
Figure 155: Home Screen with High Priority Alarm
1
Summary of Contents for MPS 3
Page 1: ...Quest Medical Inc...
Page 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Page 12: ...MPS 3 System Introduction 6...
Page 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Page 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Page 60: ...MPS 3 System System Installation 18...
Page 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Page 90: ...MPS 3 System Startup This page intentionally left blank 9...
Page 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Page 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Page 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Page 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Page 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Page 178: ...MPS 3 System Functional Use 32...
Page 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Page 231: ...MPS 3 System Troubleshooting 40...
Page 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Page 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Page 234: ...MPS 3 System Troubleshooting 43...
Page 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Page 252: ...MPS 3 System Service 4...
Page 256: ...MPS 3 System Part Numbers and Accessories 4...
Page 258: ...MPS 3 System Part Numbers and Accessories 6...
Page 259: ...MPS 3 System Part Numbers and Accessories 7...
Page 260: ......
Page 268: ...MPS 3 System Appendix This page intentionally left blank 8...