MPS 3 System – Disposable Installation
8.1.5. Installing and Filling the Additive Cartridge
1. Ensure the plunger tip is fully inserted to the end of the Arrest Agent Cartridge
barrel by pushing the plunger into the cartridge.
Figure 94: Fully Insert Plunger in Cartridge
2. Remove the plunger by unscrewing it in a counterclockwise motion
Figure 95: Plunger Removal
3. With printing facing to the right, align the two flanges on the Additive Cartridge
with the matching features on the Additive Pump on the MPS 3 Console.
DO NOT OVERFILL OR OVER PRESSURIZE arrest agent or
additive cartridges. The normal maximum cartridge capacity
is 75 ml. Overfilling the arrest agent or additive cartridges
before or after installation in the console can cause
unintentional bolus delivery and possible system
malfunction.
DO NOT REMOVE additive cartridge while delivering
cardioplegia. Removing the additive cartridge while
delivering cardioplegia can cause inaccurate delivery
calculations.
11
Cartridge Luer
Caps
Summary of Contents for MPS 3
Page 1: ...Quest Medical Inc...
Page 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Page 12: ...MPS 3 System Introduction 6...
Page 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Page 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Page 60: ...MPS 3 System System Installation 18...
Page 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Page 90: ...MPS 3 System Startup This page intentionally left blank 9...
Page 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Page 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Page 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Page 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Page 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Page 178: ...MPS 3 System Functional Use 32...
Page 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Page 231: ...MPS 3 System Troubleshooting 40...
Page 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Page 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Page 234: ...MPS 3 System Troubleshooting 43...
Page 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Page 252: ...MPS 3 System Service 4...
Page 256: ...MPS 3 System Part Numbers and Accessories 4...
Page 258: ...MPS 3 System Part Numbers and Accessories 6...
Page 259: ...MPS 3 System Part Numbers and Accessories 7...
Page 260: ......
Page 268: ...MPS 3 System Appendix This page intentionally left blank 8...