MPS 3 System – Graphical User Interface (GUI)
Overview
7.4.1. ECG Events
Selecting the ECG chart displays the ECG Event screen. An ECG cable does not
have to be connected to the Controller to enter the ECG Event screen or to record
events.
•
XClamp Time:
The time elapsed between selecting XC-ON button and
selecting XC-OFF button.
•
Time to Arrest:
The value of On Time when the Arrested button is selected.
•
Arrest Time:
The time elapsed from when selecting Arrested button to
selecting Rhythm button.
•
Ischemic Time:
The Arrest Time minus the Delivery Time.
•
Time to Rhythm:
The time elapsed between selecting XC-OFF button or
confirming Hotshot and selecting Rhythm button.
The XClamp Time can be started by selecting the XC-ON button. Select the
Arrested button to freeze and record the Time to Arrest value. The Arrest Time
counter then starts counting and the Ischemic counter starts to display the
accumulated ‘Off Time.’ When the Rhythm button is selected the operator can select
confirm to then record whether the heart was defibrillated, paced, temporarily paced
or spontaneously restarted. To enable the Rhythm button, the operator must first
select XC-OFF or confirm the Hotshot Detection when Delivery Temperature
reaches 35°C. After the Rhythm button has been confirmed, the Time to Arrest timer
resets and the operator has the option to record all of the ECG Event timers for four
additional arrest events. After the Rhythm button has been confirmed for a fifth time,
all of the buttons on the ECG Event screen become disabled.
Figure 67: ECG Event View
25
Summary of Contents for MPS 3
Page 1: ...Quest Medical Inc...
Page 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Page 12: ...MPS 3 System Introduction 6...
Page 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Page 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Page 60: ...MPS 3 System System Installation 18...
Page 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Page 90: ...MPS 3 System Startup This page intentionally left blank 9...
Page 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Page 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Page 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Page 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Page 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Page 178: ...MPS 3 System Functional Use 32...
Page 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Page 231: ...MPS 3 System Troubleshooting 40...
Page 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Page 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Page 234: ...MPS 3 System Troubleshooting 43...
Page 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Page 252: ...MPS 3 System Service 4...
Page 256: ...MPS 3 System Part Numbers and Accessories 4...
Page 258: ...MPS 3 System Part Numbers and Accessories 6...
Page 259: ...MPS 3 System Part Numbers and Accessories 7...
Page 260: ......
Page 268: ...MPS 3 System Appendix This page intentionally left blank 8...