TROUBLESHOOTING 38
CHAPTER 6
TROUBLESHOOTING
Symptom, Likely Problem and Solution
If you should experience any difficulty when using your BIOLOG 3000
i
,
please review the suggestions for troubleshooting some minor difficulties
presented below. If you cannot resolve the difficulty after following these
procedures, please contact your supplier, or call QRS Diagnostic, LLC.
No power: the unit is off after being on.
When not engaged in real-time monitoring, the BIOLOG 3000
i
will
automatically shut off after the period of time specified, (section 2.6.3
power down), to conserve battery power. Move the ON/OFF Switch to
the OFF position, then turn the unit ON again.
If the unit shuts down during download to PC or printing, it is likely that
the batteries are flat or are not of the recommended type. Replace the
batteries if this is occurring. Turning off the backlight may give some
extra operational time if no fresh batteries are immediately available.
NOTE:
Data stored in memory will not be lost.
No power: the unit will not turn on.
The batteries are flat or they are not properly installed. Remove the
Battery Compartment cover and check the position of the batteries (refer
to section 2.1.1 Installing the batteries).
Try adjusting the contrast knob. If misadjusted, no display will be visible.
If the unit is more than 4 years old, the unit may need servicing. Ensure
that the previous steps have been taken, and if necessary return it to an
authorized QRS Diagnostic, LLC service center.
'BATT' Message Appears in the Display.
Battery voltage has dropped below an acceptable level. The
BATT
BATT
message will remain on the screen until the batteries have been
replaced. Complete the procedure and change the batteries as soon as
possible.