QLS 2000/3000 Series Digital Color Printers
Printer Setup
17
The printer’s internal label queue is reset during calibration. Any labels pending
in the queue are terminated.
TO CALIBRATE THE PRINTER:
1. Press the Pause/Print button to pause the printer.
2. Press and hold the Form Feed/Advance pushbutton until the gap, notch, or
index mark separating labels on the media roll is positioned within two
inches of the media travel sensor housing.
Figure 12, Media Travel Sensor
3. Confirm that the gap, notch, string hole, or punchout is aligned with the
pointer on top of the media gap sensor: If the media has a gap between
labels, the pointer must be positioned within the width of the label. To
adjust the pointer, slide it into alignment.
1 Pointer
2 Red
slider
Figure 13, Media Sensor Assembly