Power supply

Battery powered mode: 6VDC 4×AA batteries

AC adaptor powered mode: 6V       1A (not included)

(Please only use the recommended AC adaptor model).

Display mode

Digital LCD V.A.78mm*92mm

Measurement mode

Oscillographic testing mode

Measurement range

Rated cuff pressure: 0mmHg~299mmHg(0kPa ~ 39.9kPa)

Measurement pressure:

SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)

DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)

Pulse value: (40-199)beat/minute

Accuracy

Pressure: 50C-400Cwithin±3mmHg(0.4kPa)     Pulse value:±5%

Normal working condition

A temperature range of :+5°C to +40°C

A relative humidity range of 15% to 90%,

non-condensing, but not requiring a water vapour

partial pressure greater than 50 hPa

An atmospheric pressure range of : 700 hPa to 1060 hPa

Storage & 

transportation condition

Temperature:-20°C to +60°C

A relative humidity range of ≤ 93%, non-condensing, at a water 

vapour pressure up to 50hPa

Measurement  perimeter

of the upper arm

About 22cm~42cm

Weight

Approx.393g(Excluding the batteries)

External dimensions

Approx.154.3mm*121.5mm*68.1mm

Attachment

4×AA batteries,user manual

Mode of operation

Continuous operation

Degree of protection

Type BF applied part

Protection against

ingress of water

IP21 It means the device could protected against solid foreign 

objects of 12.5mm and greater, and protect against vertically 

falling water drops.

Device Classification

Battery Powered Mode: Internally Powered ME Equipment

AC Adaptor Powered Mode: Class II ME Equipment

Software Version

A01

Problem

SYMPTOM

CHECK THIS

REMEDY

No 

power

or Low

batteries

Display will not 

light up 

or

show

Batteries are exhausted.

Replace with new batteries

Batteries are inserted

incorrectly.

Insert the batteries correctly

AC adaptor is inserted

incorrectly.

Insert the AC adaptor tightly

Error

message

E 1 shows

The cuff is not secure

or inflation is abnormal.

Refasten the cuff and then

measure again.

E 2 shows

The monitor detected

motion, talking or the

pluse is not readable

while measuring.

Movement can affect the

measurement. Relax for a

moment and then measure 

again.

E3 shows

The measurement

process does not detect

the pulse signal.

Loosen the clothing on the

arm and then measure again

E4 shows

The treatment of the

measurement failed.

Relax for a moment and

then measure again.

E5 shows

Failed to communicate

with the server

Contact customer service

E6 shows

Radio communication

failure

Contact customer service

EExx,shows on

the display.

A calibration error

occurred.

Retake the measurement.

If the problem persists, contact 

the retailer or our customer 

service department for further

assistance.Refer to the warranty 

for contact information and 

return instructions

Warning

message

‘out’ shows

Out of measurement

range

Relax for a moment. Refasten 

the cuff and then measure 

again. If the problem persists, 

contact your physician.

This section includes a list of error messages and frequently asked questions for problems you may 

encounter with your blood pressure monitor. If the device is not operating as you think it should, 

check here before arranging for servicing.

WARNING: No modification of this equipment is allowed.

TROUBLESHOOTING

SPECIFICATIONS

19

18

Summary of Contents for PY-802-LTE

Page 1: ...LTE Blood Pressure Monitor PY 802 LTE Thank you for choosing Pylo Health This manual outlines how to use your Pylo blood pressure monitor safely and correctly Please be sure to read it thoroughly and...

Page 2: ...Principle Safety Information LCD Display Signal Components of the Device Included Parts List What are systolic pressure and diastolic pressure What is the standard blood pressure classification Irreg...

Page 3: ...on and provides step by step instructions for using the product Please read this manual thoroughly before using the product This product uses the Oscillometric Measuring Method to detect blood pressur...

Page 4: ...eck that operation of the device does not result in prolonged impairment of patient blood circulation When measurement please avoid compression or restriction of the connection tubing The device canno...

Page 5: ...he device doesn t need to be calibrated within two years of reliable service If you have any problems with this device such as setting up maintaining or using please contact the SERVICE PERSONNEL of P...

Page 6: ...d special power adapter which complies with local safety standard Do not use new and used batteries together Do not use different types of batteries together Do not dispose the batteries in fire Batte...

Page 7: ...in place Don t make it too tight allow a finger to fit between the cuff and your arm 10 If possible relax and rest for at least 5 minutes before taking a measurement 11 Lay your arm on a table with y...

Page 8: ...urn on the monitor The cuff will begin to inflate and the measurement will be taken and saved 2 Once measurement is taken the device will connect to the network and start transmission The symbol will...

Page 9: ...e device In a very cold environment When cleaning device use a lightly moist cloth without soap or detergent After drinking tea coffee or using tobacco products Keep away from water If splashed immedi...

Page 10: ...he person takes medicine the pressure may vary even more 3 Wait at least 3 minutes between measurements What you need to pay attention to when you measure your blood pressure at home If the cuff is ti...

Page 11: ...1 Problem SYMPTOM CHECK THIS REMEDY No power or Low batteries Display will not light up or show Batteries are exhausted Replace with new batteries Batteries are inserted incorrectly Insert the batteri...

Page 12: ...to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential ins...

Page 13: ...rer of medical devices General Requirements for Safety EN 60601 1 2006 A1 2013 IEC 60601 1 2005 A1 2012 Medical electrical equipment Part 1 General requirements for basic safety and essential performa...

Page 14: ...z Conduced RF IEC61000 4 6 3 V 0 15 MHz 80 MHz 6 V in ISM and amateur radio bands between 0 15 MHz and 80 MHz 80 AM at 1 kHz 3 V 0 15 MHz 80 MHz 6 V in ISM and amateur radio bands between 0 15 MHz and...

Reviews: