Power supply
Battery powered mode: 6VDC 4×AA batteries
AC adaptor powered mode: 6V 1A (not included)
(Please only use the recommended AC adaptor model).
Display mode
Digital LCD V.A.78mm*92mm
Measurement mode
Oscillographic testing mode
Measurement range
Rated cuff pressure: 0mmHg~299mmHg(0kPa ~ 39.9kPa)
Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute
Accuracy
Pressure: 50C-400Cwithin±3mmHg(0.4kPa) Pulse value:±5%
Normal working condition
A temperature range of :+5°C to +40°C
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water vapour
partial pressure greater than 50 hPa
An atmospheric pressure range of : 700 hPa to 1060 hPa
Storage &
transportation condition
Temperature:-20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing, at a water
vapour pressure up to 50hPa
Measurement perimeter
of the upper arm
About 22cm~42cm
Weight
Approx.393g(Excluding the batteries)
External dimensions
Approx.154.3mm*121.5mm*68.1mm
Attachment
4×AA batteries,user manual
Mode of operation
Continuous operation
Degree of protection
Type BF applied part
Protection against
ingress of water
IP21 It means the device could protected against solid foreign
objects of 12.5mm and greater, and protect against vertically
falling water drops.
Device Classification
Battery Powered Mode: Internally Powered ME Equipment
AC Adaptor Powered Mode: Class II ME Equipment
Software Version
A01
Problem
SYMPTOM
CHECK THIS
REMEDY
No
power
or Low
batteries
Display will not
light up
or
show
Batteries are exhausted.
Replace with new batteries
Batteries are inserted
incorrectly.
Insert the batteries correctly
AC adaptor is inserted
incorrectly.
Insert the AC adaptor tightly
Error
message
E 1 shows
The cuff is not secure
or inflation is abnormal.
Refasten the cuff and then
measure again.
E 2 shows
The monitor detected
motion, talking or the
pluse is not readable
while measuring.
Movement can affect the
measurement. Relax for a
moment and then measure
again.
E3 shows
The measurement
process does not detect
the pulse signal.
Loosen the clothing on the
arm and then measure again
E4 shows
The treatment of the
measurement failed.
Relax for a moment and
then measure again.
E5 shows
Failed to communicate
with the server
Contact customer service
E6 shows
Radio communication
failure
Contact customer service
EExx,shows on
the display.
A calibration error
occurred.
Retake the measurement.
If the problem persists, contact
the retailer or our customer
service department for further
assistance.Refer to the warranty
for contact information and
return instructions
Warning
message
‘out’ shows
Out of measurement
range
Relax for a moment. Refasten
the cuff and then measure
again. If the problem persists,
contact your physician.
This section includes a list of error messages and frequently asked questions for problems you may
encounter with your blood pressure monitor. If the device is not operating as you think it should,
check here before arranging for servicing.
WARNING: No modification of this equipment is allowed.
TROUBLESHOOTING
SPECIFICATIONS
19