Page 230
•
Keys
. You should press each key, to confirm it works, with a
counter showing how many more keys you have to press. Be sure
to press the
CANCEL
key last, as this will show if all keys were
pressed or not. If they were not, then an error message is displayed.
P992 mA Out Test
This parameter will allow you to force a specified current on the mA output,
to test the equipment that it is connected to, and to make sure the unit is
working correctly. The figure you enter will be generated by the mA output.
P993 mA In Test (Optional)
This parameter will allow you to test the mA input, by injecting a known
mA signal from an external source you can check the unit is working
correctly and as expected.
P994 Transducer Test
If you enter 1 for this parameter it will continually fire the transducer, so
you can check the wiring, until you press any key to cancel.
P995 Keys Test
You should press each key, to confirm it works, with a counter showing how
many more keys you have to press. Press the
CANCEL
key last, as this will
confirm if all keys were pressed or not. If they were not, then an error
message is displayed.
P996 Relay Test
Press the numeric key corresponding to the number of the relay you wish to
test, and the relay will change state each time the key is pressed. If you press
any other key, other than a valid relay number, then the test will end.
Summary of Contents for ULTRA 5
Page 1: ...ULTRA 5 INSTRUCTION MANUAL ...
Page 2: ......
Page 4: ......
Page 15: ...Page 7 EU Declaration of Conformity Wall Mount ...
Page 16: ...Page 8 Fascia Mount ...
Page 17: ...Page 9 Rack and Panel ...
Page 18: ...Page 10 This page left blank intentionally ...
Page 21: ...Page 13 The full dimensions of the enclosure are as shown below ...
Page 23: ...Page 15 The full dimensions of the Fascia mount enclosure are as shown below ...
Page 35: ...Page 27 This page left blank intentionally ...
Page 151: ...Page 143 This page left blank intentionally ...
Page 240: ...Page 232 This page left blank intentionally ...