PT610, PT620, PT630
Instruction Manual
62
6.1 ENTERING AND EXITING SETUP
When entering set up, please be reminded of the key functions as below. Refer to the buttons for the
appropriate indicator described in 5.1.1
<BACK>
<DOWN>
<UP>
<CHANGE>
<ENTER>
Return.
Decrease or
previous step.
Increase or
next step.
Change the
item or enter
into the
parameter.
Confirm the
entry or
selection and
go to the next
step.
A set-up DIP switch is located on the main PCB near the load cell connector, as shown in the figure
below and its position should be “ON” to change the metrological related parameters including
calibration.
Set position 1 to ON for calibration, leave switch position 2, it is reserved.
Entering Setup - Access at the start
•
To access the metrological parameters in grey in the menu above set the DIP switch deeply
recessed on the back panel into the down position. (See 4.8)
Display
Operation
[
123.456
kg]
Press and hold the
<ENTER>
key until the
[PASSWORD:]
prompt
is seen.
[PASSWORD:]
For the PT610, press the
<X10> <ZERO>
<TARE>
<ZERO>
keys
sequentially in order. For the PT620 and PT630 enter the 4 digit
passcode with the numerical keypad (default 1111).
[**** ]
Press the
<ENTER>
key to confirm.
[1—INTERFACE ]
The first block of the programming menu is displayed.
Setup - Fast Access
•
To quickly jump to various areas of the menu, press and hold the suitable key listed below for
more than 2 seconds when in the main block
[1—INTERFACE ]
.
Fast Access Key
Function
<FASTOPT>
Fast access to option card parameters
[19-]
(if fitted).
<FASTCAL>
Fast access to calibration
[621]
.
<STOR>
Fast access to EZAZ parameters (not used on PT610).
<FASTSET>
Fast access to service parameters.