About setpoint and tolerances
B0232
P
P
P
P
P
P
P
calibration
stroke feedback
mass (stroke)
0.347 g
stroke feedback
set-point
mass (stroke)
0.347 g
lower tolerance
-010.00 %
tolerance
tolerance (%)
upper tolerance
+020.00 %
Specify the permitted range for the stroke mass m/H via the setpoint of the stroke mass m/H and
tolerances in %:
1.
Follow the
‘calibration
è
stroke feedback
è
setpoint’ menu path and press [P].
ð
The
‘mass (stroke)’ menu item displays the setpoint currently saved.
2.
Start the metering pump.
ð
The current measured value is displayed.
3.
Turn the stroke length adjustment knob until the required setpoint is displayed.
4.
Press
[P].
ð
The measured value displayed is saved as a setpoint and the
‘tolerance tolerances’
menu item appears.
The setpoint is valid as 100 %.
5.
Press
[P].
6.
Use the
[arrow keys] to enter the ‘lower tolerance’ and press [P].
7.
Use the
[arrow keys] to enter the ‘upper tolerance’ and press [P].
ð
The continuous display appears again.
8.
Stop the metering pump.
The required setpoint can also be entered without the need to use the stroke adjustment dial
or the pump to be running, by using the [arrow keys] directly in ‘mass (stroke)’ .
Set up