Press
[ i ] (“i” for “Information”) if the
LCD screen displays a continuous dis‐
play - i.e. does not display the operating
menu.
ð
Another continuous display
appears every time you press the
[i]
key.
The appearance of the continuous displays
depends on the measured variable selected
(volumes or mass) - see below.
Press [ i ] to move from continuous dis‐
play to continuous display.
Continuous displays for the “Volume” measured
variable
Continuous display
Description
V:
243,32 mL
Q:
171,05 mL/h
Total volume* V and
current volumetric
flow Q
V:
243,32 mL
N:
637
Total volume* V and
number of strokes* N
V/H:
0,382 mL
%SH: 102,3 %
Volume per stroke
V/H and deviation
from the stroke
volume setpoint %SH
* since the last reset
Continuous displays for the “Mass” measured
variable
(If
‘mass’ was set under ‘Mode
è
Measured variable’).
Continuous display
Description
m:
326,05 g
Q:
229,21 g/h
Total volume* m and
current mass flow Q
m:
326,05 g
N:
527
Total volume* m and
number of strokes* N
m/H:
0,619 g
%SH: 97,8 %
Volume per stroke
m/H and deviation
from the stroke mass
setpoint %SH
* since the last reset
Resetting values
–
Press the two [arrow keys] simul‐
taneously to reset the total volume
and strokes.
–
You can reset the total volume and
the strokes independently of each
other in the “zero set” main menu.
Set up
29