PIN Number and Calibration Switch
_____________________________________________________________________________________________________________________________________
6-4
6.6 PIN Number and Calibration Switch
Calibration can be protected against accidental or incorrect calibrations, through the
use of a rear calibration switch and a special PIN number.
The calibration button is located on the rear panel, and can be pushed with a pointed
object (pen, pencil, paper clip, etc.). This button can be pushed after the calibration
menu has been selected. At the same time, a request to input the PIN number will
appear in the display.
Press Key
Display / Action
1.
Cal key on the
rear panel
1 P I N : 0 0 0 0 0 0 0
Enter a PIN No. (after delivery: 0000000), the display shows alternating
"CAL" and the reading with the correct PIN No..
2.
MENU
-
IN KEY
2 C O N F I G U R E . .
if this display does not appear press the
ñ
KEY
until it appears
3.
MENU
-
IN KEY
4 C a l i b r a t i o n . .
if this display does not appear press the
ñ
KEY
until it appears
4.
MENU
-
IN KEY
1 E n t e r V a l u e . .
only with the right PIN No..
5.
MENU
-
IN KEY
V + 1 0 0 . 0 0 0 0 0
Enter the calibration value, close with the
MENU
-
IN KEY
6.
M
ENU
-
IN KEY
1 E n t e r V a l u e . .
press several times to return to the calibration input menu
7.
ò
KEY
2 S t o r e C a l .
with afterwards pressing of the
MENU
-
IN KEY
stores durably the
calibration data
8.
Cal key on the
rear panel
to close the calibration procedure
Note:
Item 7 is very important. If you forget to store the user calibration it is
completely deleted after you switch off the instrument. Then only the
factory calibration is valid.