9 Technical Specifications
_____________________________________________________________________________________________________________________________________
9-9
OVERLOAD LIMITS
V/
Ω
Ω
-HI against Case
± 1000 V Peak at 60 Hz max.
or ± 1000V DC Voltage
V/
Ω
Ω
-HI against V/
Ω
Ω
-LO
200mV-, 2V-Range
for 60 sec.
± 1000V Peak with the
limit of 1*10
7
*V*Hz
Continuous Load
± 700V Peak with the
limit of 5*10
6
*V*Hz
20V-, 200V-, 700V-Range 3) Continuous Load ± 1000V Peak with the
limit of 1*10
7
*V*Hz
with built-in Scanner (Model 5017SC)
Continuous Load on rear panel connectors
± 125V Peak with the
limit of 2*10
5
*V*Hz
V/
Ω
Ω
/LO against Ground
250 Vdc or 250 Vpk
MEASUREMENT PAUSES
after Range, Function, or Channel Switching:
200mV
300ms
2V, 20V
200ms
200V, 700V
250ms
CREST FACTOR
max. 7:1
INPUT RESISTANCE
1 M
Ω
in parallel >60pF
SETTLING TIME
100ms at 0.1%
1) V/
Ω
/LO connected to Ground, Sine Signal larger than 5% of Full Scale
2) % F.S. is always related to a Readout Span of 20 100 000