Resistance
_____________________________________________________________________________________________________________________________________
9-6
9.2 Resistance
MEASURING METHOD
.....
2-, 4-pole with DC measuring current
RANGES
...............................
300
Ω
/ 3 k
Ω
/ 30 k
Ω
/ 300 k
Ω /
3 M
Ω
/ 30 M
Ω
RANGE SELECTION
...........
manual, automatic
MEASUREMENT TIMES
Full Scale
max. Resolution
20ms / 40ms / 100ms
301 000
1 m
Ω
0.2s / 0.4s / 1s
3 010 000
100 µ
Ω
2 / 4 / 10 / 20 / 40 / 100 s 30 100 000
10 µ
Ω
STABILITY
24 Hours, 23°C ± 1°C
1), 2)
± (% of Reading + % of Full Scale)
Range
% Rdg.
% F.S.
300
Ω
0.001
0.0003
3k
Ω
0.0004
0.0002
30k
Ω
0.0004
0.0002
300k
Ω
0.0006
0.0002
3M
Ω
0.004
0.0003
30M
Ω
0.020
0.0010
ACCURACY
(1 Year, 23°C±5°C)
1), 2)
± (% of Reading + % of Full Scale)
Range
% Rdg.
% F.S.
300
Ω
0.003
0.0005
3k
Ω
0.003
0.0004
30k
Ω
0.003
0.0004
300k
Ω
0.003
0.0004
3M
Ω
0.007
0.0004
30M
Ω
0.050
0.0010