5 Remote Control
_____________________________________________________________________________________________________________________________________
5-13
Tx
Integration time / Measuring time selection with
x = 0
20ms
x = 6
2 s
x = 1
40ms
x = 7
4 s
x = 2
100ms
x = 8
10 s
x = 3
200ms
x = 9
20 s
x = 4
400ms
x = A
40 s
x = 5
1 s
x = B
100 s
x = Z
10ms (only for frequency measurement)
D0
(D/zero) Display mode switch-off
D1“text“
Display mode switch-on. A text sent after “D1“ is output to the dis
play of the multimeter. The internal display is switched off. The
text
sent must be enclosed between quote characters.
F0
(F/zero) switches off the additional filter.
F1
Switches on the average value filter (Avg. Filter).
F2
Switches on the automatic filter (Auto Filter ).
F3
Switches on the fast automatic filter (Fast Auto Filter ).
Q0
(Q/zero) switches off SRQ mode.
Q1
Switches on SRQ mode on:
- every new measurement reading
- error message
- reset
- out of limit
EQ?
Query of the Error-Queue
S0
(S/zero) start mode off, continuous sequence of measurements on
S1
Start mode on; every S1 command starts a measurement
S2
Start mode on, start with trigger line or trigger button