Measuring Voltage
_____________________________________________________________________________________________________________________________________
3-2
3.3 Measuring Voltage
Measurement Functions
DC Voltage
VDC
key
AC Voltage
VAC
key
AC Voltage with DC Component
VAC
key and
COUPL
key
Measurement Ranges
Vdc
300mV, 3V, 30V, 300V, 1000V
Vac
200mV, 2V, 20V, 200V, 700V
Resolution and Measurement Times
Vdc
20 / 40 / 100 ms
5½ digits
200 / 400 ms / 1s
6½ digits
2 / 4 / 10 / 20 / 40 / 100 s
7½ digits
Vac
100 ms
5½ digits
0.2 / 0.4 / 1 / 2 / 4 / 10 / 20 / 40 / 100s
6½ digits
max. Resolution in the smallest range
Vdc: 10nV Vac: 100nV
Frequency Range
Vac: 20 Hz to 1 MHz
Figure: Connection of measurement leads in Voltage Measurement
The
rear plugs
are connected correspondingly.
For model 5017SC with scanner the measurement leads are connected to Hi and Lo of
the respective channel.
HI
HI
Amps
V
Ω Ω4
Sense
LO
+
__
DC or AC Voltage