The connector terminal may be un-plugged from the circuit board
to simplify installation of the remote signal wires.
Land the corresponding signal on the NC (normally closed) &
common wires on the C (common) terminal of each pair of terminal
connectors. It is imperative that the signal wires are landed in the NC
position and that the common wires are landed in the C position.
Wiring the Remote Devices to Remote Signal Master Board
(Transducer Sensors)
Locate the bank of 8 terminal connectors on the left edge of the
back of the remote signal (master) board.
There are a total of 8 connection points – two connection points
for each transducer sensor. The pairs are labeled 1 IN thru 4 IN on
the circuit board.
The white wire from the transducer should be
landed in the WHT position and the black wire from the transducer
should be landed in the BLK position. Note: only Powerex D and T
series alarm transducers are compatible with the Med Touch alarm
system.
Remote Signal #1
Common Wire Signal #1
Remote Signal #2
Common Wire Signal #2
The connector terminal may be un-plugged from the circuit board
to simplify installation of the remote signal wires.
The sequence in which the transducers are connected to the board
correspond with the order in which they will appear on the display.
The #1 transducer will be displayed on top and the #4 transducer will
be displayed on the bottom. There will be up to two display positions
(high line pressure and low line pressure) for each transducer.
Transducer #1 White
Transducer #1 Black
Transducer #2 White
Transducer #2 Black