UL CLASSIFICATION 

 
The Porter Digital MXR-D is UL Classified 82AN:  Medical Equipment with respect to electric 
shock, fire and mechanical hazard only in accordance with UL 60601-1 and CAN/CSA C22.2 No. 
601.1.UL. The system consists of the Digital MXR-D and Power Supply.  

 

The following are UL classifications for the Porter Digital MXR-D: 

 

1.  Is not intended to be used with or in the presence of flammable anesthetic mixtures. 

2.  Is classified as ordinary equipment and is designed for continuous operation. 

3.  Has safety protection provided in the equipment.  If the equipment is tampered with or 

used in a manner not specified in this manual the safety protection provided by the 
equipment might be impaired. 

 

The Porter Digital MXR-D is designed to be used in normal environmental conditions: 

1. Indoor 

use 

2.  Altitude up to 8000 feet (2,438 meters) 

3.  Temperature from 5°C to 40°C 

4.  Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 

50% relative humidity at 40°C 

5.  Mains supply voltage fluctuations up to +/- 10% of the nominal voltage 

6.  Transient over voltages typically present on mains supply 

7.  Normal pollution degree 

 
The Porter Digital MXR-D controls both Oxygen (O

2

) and Nitrous Oxide (N

2

O).  Both gases could 

be present in an over pressure event that requires the manifold system to have a safety relief 
valve to function.  
 
Maximum system pressure is not to exceed 60 PSIG. 
 
Product was tested for Electromagnetic Immunity per IEC 61000-4-3. 

 

 

5

Summary of Contents for 4065D

Page 1: ...DIGITAL MXR D Cabinet Mount Flowmeter Component Diagram and Installation Instructions MODEL 4065D FM 1062 Rev B 6 07...

Page 2: ...CSA C22 2 No 601 1 M90 Symbol Description Protective Earth Ground Class I Device Alternating Current Earth Ground Attention consult the accompanying documents Type B Equipment Distribution of Porter M...

Page 3: ...s 9 NFPA Codes 9 INTENDED USE Flowmeter for Analgesia Delivery System 9 INSTALL THE GAS LINES 10 SWITCHING THE UNIVERSAL CONNECTORS 10 CABINET WALL MOUNT INSTALLATION 11 FLOWMETER ACCESSORIES 13 CHECK...

Page 4: ...is to be wired per NEC and local electrical codes The branch circuit breaker is the main disconnect Porter recommends the following 15A UL circuit breaker two wire with ground TANK REGULATOR SPECIFICA...

Page 5: ...anual the safety protection provided by the equipment might be impaired The Porter Digital MXR D is designed to be used in normal environmental conditions 1 Indoor use 2 Altitude up to 8000 feet 2 438...

Page 6: ...904 or call 1 800 344 3555 WARNING New or modified installations properly connected gas pipelines are absolutely essential to patient safety The dealer or contractor should provide written documentati...

Page 7: ...nd alarms needed for the Vanguard test Then turn off both Oxygen tanks and leave both Nitrous Oxide tanks on For the Sentinel only one tank of each gas is on at a time Turn off the one open Oxygen tan...

Page 8: ...outlet station the Oxygen 1 2 inch pipe is incorrectly decreased down to the DISS 3 8 inch size reserved for Nitrous Oxide lines Oxygen is fed into the Nitrous Oxide lines At gas outlet station the c...

Page 9: ...Analgesia Delivery System This device is intended for patient adult and pediatric use by an attending physician or dentist properly trained in its use Porter Instrument recommends the user be thorough...

Page 10: ...ter The dental dealer is responsible for the final connections to the flowmeter the details of flowmeter installation checking for crossed lines leak testing of connections to the flowmeter and basic...

Page 11: ...ce a piece of tape over the rear power supply connector temporarily so dry wall dust does not enter the connector Wrenches required 7 8 for N2O connector 5 8 for O2 connector and 7 16 for mixed gas co...

Page 12: ...e O2 shut off valve and the N2O DISS hose blue to the N2O shut off valve 9 Run 20 feet mixed gas hose from flowmeter mixed gas outlet hose barb to hose barb of Remote Surface Mount or Through Wall Ada...

Page 13: ...ff position Confirm that the DISS Shut Off Valves are in the open position Pressurize the sedation gas supply lines with 50 PSI Observe any pressure decay after an overnight time period 5 PSI drop all...

Page 14: ...POWER SUPPLY SYSTEM ILLUSTRATION 14...

Page 15: ...GAS DISTRIBUTION SYSTEM ILLUSTRATION 15...

Page 16: ...w Knobs to full mechanical stops Installation Check for proper display STEP 1 N2O L M There is no Oxygen flowing There is no Nitrous flowing N2O and Total Flow valves are shut tightly Normal display c...

Reviews: