Back Connection Installation:   

a.)  Remove red caps from connector. 
b.)  Connect gas piping to meter and press 
unit into mounting frame.  

 

Bottom Connection Installation:   

 

a.) Follow the instructions “Switching the 
Universal Connectors” in this manual to 
remove the N

2

O, O

2

 and Mixed Gas 

connectors and switch to the bottom 
position.  After removing the connectors, 
press unit into mounting frame. 
b.) Re-install connectors in the bottom 
position following instructions for “Switching 
the Universal Connectors”. 
c.) Remove red caps from connectors. 
d.) Connect gas piping to meter. 

 
5. 

Connect power supply to flowmeter. 

IMPORTANT:

  Connector from power supply 

to flowmeter must be fully pressed into 
mating connector to ensure that locking 
feature is engaged. Refer to “Power Supply 
System Illustration”. 

6.  Connect Porter supplied “Hospital Grade” 

power cord. 

7.  Determine the location of remote bag tee 

that accommodates visual monitoring of 
breathing bag during patient treatment. Drill 
a 3/4” hole in the cabinet that will 
accommodate the Through Wall Adapter 
Assembly.  Alternatively, use the Remote 
Surface Mount.   

Note:

 Bag ideally is to be 

installed within 10 feet of the flowmeter. 

8.  Connect the O

2

 DISS hose (green) to the O

2

 

shut-off valve and the N

2

O DISS hose (blue) 

to the N

2

O shut-off valve.  

 

 

9.  Run 20 feet mixed gas hose from flowmeter 

mixed gas outlet hose barb to hose barb of 
Remote Surface Mount or Through Wall 
Adapter Assembly. Cut off any excess hose. 

10. Install remote bag tee. Screw knurled seal 

nut down tight onto remote adapter, making 
sure the rubber washer is inside the seal nut.  
Bag tee should not rotate. Refer to FM-809 
for User Instructions for Assembly, 
Installation and Cleaning of Porter 
Scavenger Rubber Goods System.  

Note:

 

Bag Tee ideally is to be installed within 10 
feet of the flowmeter. 

 

Optional 
Bezel Kit 
B-4988

 

11. Pull  O

2

 and N

2

O DISS hoses and mixed gas 

hose through opening in mounting frame. 

12. Hold cabinet mount flowmeter up to the 

mounting frame and attach O

2

 and N

2

DISS hoses to Universal connectors of 
flowmeter.  

 

Caution:

 Make certain hoses are not 

crimped.  Always 

use two wrenches when 

tightening

, to assure that the Universal 

connectors are not loosened from the 
flowmeter cavities.  Verify that the Universal 
connectors are tightened firmly to the 
flowmeter and that the DISS hoses are 
tightened firmly to the Universal connectors. 

 

13. Place the mixed gas hose on the flowmeter 

mixed gas outlet hose barb. 

14. Attach retaining spring to meter body using 

two screws.  Repeat for opposite side. 

15. Seat flowmeter into position by pressing it 

firmly into the mounting frame.  

 

12

Summary of Contents for 4065D

Page 1: ...DIGITAL MXR D Cabinet Mount Flowmeter Component Diagram and Installation Instructions MODEL 4065D FM 1062 Rev B 6 07...

Page 2: ...CSA C22 2 No 601 1 M90 Symbol Description Protective Earth Ground Class I Device Alternating Current Earth Ground Attention consult the accompanying documents Type B Equipment Distribution of Porter M...

Page 3: ...s 9 NFPA Codes 9 INTENDED USE Flowmeter for Analgesia Delivery System 9 INSTALL THE GAS LINES 10 SWITCHING THE UNIVERSAL CONNECTORS 10 CABINET WALL MOUNT INSTALLATION 11 FLOWMETER ACCESSORIES 13 CHECK...

Page 4: ...is to be wired per NEC and local electrical codes The branch circuit breaker is the main disconnect Porter recommends the following 15A UL circuit breaker two wire with ground TANK REGULATOR SPECIFICA...

Page 5: ...anual the safety protection provided by the equipment might be impaired The Porter Digital MXR D is designed to be used in normal environmental conditions 1 Indoor use 2 Altitude up to 8000 feet 2 438...

Page 6: ...904 or call 1 800 344 3555 WARNING New or modified installations properly connected gas pipelines are absolutely essential to patient safety The dealer or contractor should provide written documentati...

Page 7: ...nd alarms needed for the Vanguard test Then turn off both Oxygen tanks and leave both Nitrous Oxide tanks on For the Sentinel only one tank of each gas is on at a time Turn off the one open Oxygen tan...

Page 8: ...outlet station the Oxygen 1 2 inch pipe is incorrectly decreased down to the DISS 3 8 inch size reserved for Nitrous Oxide lines Oxygen is fed into the Nitrous Oxide lines At gas outlet station the c...

Page 9: ...Analgesia Delivery System This device is intended for patient adult and pediatric use by an attending physician or dentist properly trained in its use Porter Instrument recommends the user be thorough...

Page 10: ...ter The dental dealer is responsible for the final connections to the flowmeter the details of flowmeter installation checking for crossed lines leak testing of connections to the flowmeter and basic...

Page 11: ...ce a piece of tape over the rear power supply connector temporarily so dry wall dust does not enter the connector Wrenches required 7 8 for N2O connector 5 8 for O2 connector and 7 16 for mixed gas co...

Page 12: ...e O2 shut off valve and the N2O DISS hose blue to the N2O shut off valve 9 Run 20 feet mixed gas hose from flowmeter mixed gas outlet hose barb to hose barb of Remote Surface Mount or Through Wall Ada...

Page 13: ...ff position Confirm that the DISS Shut Off Valves are in the open position Pressurize the sedation gas supply lines with 50 PSI Observe any pressure decay after an overnight time period 5 PSI drop all...

Page 14: ...POWER SUPPLY SYSTEM ILLUSTRATION 14...

Page 15: ...GAS DISTRIBUTION SYSTEM ILLUSTRATION 15...

Page 16: ...w Knobs to full mechanical stops Installation Check for proper display STEP 1 N2O L M There is no Oxygen flowing There is no Nitrous flowing N2O and Total Flow valves are shut tightly Normal display c...