TECHNICAL SPECIFICATIONS 

 

Physical: 

Dimensions: 12.6” W; 4.3” H; 3.9” D 
Cut out hole: 11” x 3 5/8” ;  Weight:  8 lbs. 

Fittings: 

Fresh Gas Outlet:  15mm ID x 22mm OD 
(Diameter Index Safety System) 
Nitrous Oxide Inlet:  Male DISS (CGA # 1040A) 
Oxygen Inlet:  Male DISS (CGA # 1240) (Diameter 
Index Safety System) 

Environmental: 

Storage Temperature:  -10°F to 150°F (Allow to 
stabilize to room temperature before operating.) 
Operating Temperature: 60°F to 85°F 
Humidity:  Ambient, non-condensing 

Gas Supply:  

Oxygen Inlet 
Pressure Requirements:  50 +/-10 PSIG 
Flow Requirements:  30 L/min Minimum Flow 

Note: 

 A minimum Oxygen flow of 120 L/min is 

required when using the optional Demand Valve 
Resuscitator 
Nitrous Oxide Inlet 
Pressure Requirements:  50 +/-10 PSIG 
Flow Requirements:  10 L/min Minimum Flow 

Gas Delivery: 

Oxygen Flush 35 L/min Minimum (50 PSIG input) 
Minimum Nitrous Oxide Concentration:  0% 
Maximum Nitrous Oxide Concentration:  70% 
Minimum Oxygen Concentration:  30% 
Maximum Oxygen Concentration:  100% 
Flow Range at 100% Oxygen:  1 to 10 L/min 
Total Flow Range at % Mixture: up to 12 L/min 
(Low end flow limited by 1.0 L/min Oxygen limit. 
Accuracy: 

 

± 0.5 L/M Oxygen flow and ± 0.5 L/M 

Nitrous Oxide flow.  

Electrical Power Supplies:

 

 

Wall Plug-in: 
 

Input: 115 Vac, 60 Hz, 0.1A Max 

 

(Unit is for 115 Vac use only.) 

 

Output: 12Vdc / 0.3A Max. 

 

The Branch Circuit breaker is the main 
disconnect. 

In Wall Junction Box: 
 

Input: 115 Vac, 60 Hz, 0.1A Max. 

 

Output: 12 Vdc / 0.4A Max. 

 
The main 110 Vac Branch connection is to be 
wired per NEC and local electrical codes.

  The 

branch circuit breaker is the main 

disconnect. 

Porter recommends the following:  15A UL circuit 
breaker two-wire with ground. 
 

TANK REGULATOR SPECIFICATIONS 

Oxygen Regulator 

Use Porter O

2

 Regulator part number 7000-2 

or the Oxygen tank regulator should have the 
following specifications: 

ƒ

  Regulator should be pre-set at 50 

PSIG ± 2 PSIG (static) 

ƒ

  Inlet fitting per CGA 540 

ƒ

  Outlet fitting per O

2

 DISS 

ƒ

  I0 micron filter on inlet 

ƒ

  Self-resetting safety relief valve pre-set 

to 100 ± 10 PSIG 

ƒ

  0-4000 PSIG color coded pressure 

gage 

ƒ

  Maximum flow capacity of no less than 

150 LPM 

ƒ

  Droop shall be a maximum change in 

regulated pressure of 8% from 0-150 
LPM when supply is 2000 PSIG. 

ƒ

  Regulation of the supply pressure 

should not change by more than 4% of 
set pressure at a flow of 75 LPM. 

ƒ

 UL 

Listed 

ƒ

  Installation shall conform to local codes 

with respect to pressure relief. 

 

Nitrous Oxide Regulator 

Use Porter N

2

O Regulator part number 7500-2 

or the Nitrous Oxide tank regulator should 
have the following specifications: 
 

ƒ

  Regulator should be pre-set at 50 

PSIG ± 2 PSIG (static) 

ƒ

  Inlet fitting per CGA 326 

ƒ

  Outlet fitting per N

2

O DISS 

ƒ

  I0 micron filter on inlet 

ƒ

  Self-resetting safety relief valve pre-set 

to 100 ± 10 PSIG 

ƒ

  0-1500 PSIG color coded pressure 

gage 

ƒ

  Maximum flow capacity of no less than 

100 LPM 

ƒ

  Droop shall be a maximum change in 

regulated pressure of 8% from 0-150 
LPM when supply is 800 PSIG. 

ƒ

  Regulation of the supply pressure 

should not change by more than 4% of 
set pressure at a flow of 75 LPM. 

ƒ

 UL 

Listed 

ƒ

  Installation shall conform to local codes 

with respect to pressure relief. 

 

 

4

Summary of Contents for 4065D

Page 1: ...DIGITAL MXR D Cabinet Mount Flowmeter Component Diagram and Installation Instructions MODEL 4065D FM 1062 Rev B 6 07...

Page 2: ...CSA C22 2 No 601 1 M90 Symbol Description Protective Earth Ground Class I Device Alternating Current Earth Ground Attention consult the accompanying documents Type B Equipment Distribution of Porter M...

Page 3: ...s 9 NFPA Codes 9 INTENDED USE Flowmeter for Analgesia Delivery System 9 INSTALL THE GAS LINES 10 SWITCHING THE UNIVERSAL CONNECTORS 10 CABINET WALL MOUNT INSTALLATION 11 FLOWMETER ACCESSORIES 13 CHECK...

Page 4: ...is to be wired per NEC and local electrical codes The branch circuit breaker is the main disconnect Porter recommends the following 15A UL circuit breaker two wire with ground TANK REGULATOR SPECIFICA...

Page 5: ...anual the safety protection provided by the equipment might be impaired The Porter Digital MXR D is designed to be used in normal environmental conditions 1 Indoor use 2 Altitude up to 8000 feet 2 438...

Page 6: ...904 or call 1 800 344 3555 WARNING New or modified installations properly connected gas pipelines are absolutely essential to patient safety The dealer or contractor should provide written documentati...

Page 7: ...nd alarms needed for the Vanguard test Then turn off both Oxygen tanks and leave both Nitrous Oxide tanks on For the Sentinel only one tank of each gas is on at a time Turn off the one open Oxygen tan...

Page 8: ...outlet station the Oxygen 1 2 inch pipe is incorrectly decreased down to the DISS 3 8 inch size reserved for Nitrous Oxide lines Oxygen is fed into the Nitrous Oxide lines At gas outlet station the c...

Page 9: ...Analgesia Delivery System This device is intended for patient adult and pediatric use by an attending physician or dentist properly trained in its use Porter Instrument recommends the user be thorough...

Page 10: ...ter The dental dealer is responsible for the final connections to the flowmeter the details of flowmeter installation checking for crossed lines leak testing of connections to the flowmeter and basic...

Page 11: ...ce a piece of tape over the rear power supply connector temporarily so dry wall dust does not enter the connector Wrenches required 7 8 for N2O connector 5 8 for O2 connector and 7 16 for mixed gas co...

Page 12: ...e O2 shut off valve and the N2O DISS hose blue to the N2O shut off valve 9 Run 20 feet mixed gas hose from flowmeter mixed gas outlet hose barb to hose barb of Remote Surface Mount or Through Wall Ada...

Page 13: ...ff position Confirm that the DISS Shut Off Valves are in the open position Pressurize the sedation gas supply lines with 50 PSI Observe any pressure decay after an overnight time period 5 PSI drop all...

Page 14: ...POWER SUPPLY SYSTEM ILLUSTRATION 14...

Page 15: ...GAS DISTRIBUTION SYSTEM ILLUSTRATION 15...

Page 16: ...w Knobs to full mechanical stops Installation Check for proper display STEP 1 N2O L M There is no Oxygen flowing There is no Nitrous flowing N2O and Total Flow valves are shut tightly Normal display c...