SODIMAT - Model 9073
1-3
Calibration
- Manual by 2 known concentration solutions
- Process on one point
2 versions possible :
- addition 1 or 2 points (CARCAL 20 and 200
ppb cartridges)
- automatic by dosing pump with a known
concentration solution ( 1 or 2 additions)
TRANSMITTER
Display
4 digit digital LCD , 18 mm-height, backlit
Display of the measurement unit : ppb, ppm,
mV, C or l/h, LED and the instrument status
B
: error, calibration, programming.
Output current
2 analogue outputs : 2 x 0/4... 20 mA isolated
from input signal, 900 Ohms load maximum
Relays
2 isolated contacts, minimum or maximum
switchable
1 isolated contact for system alarm
250 VAC max., 5A max. or 1200 VA on a
resistive load
Temporisation
On the concentration limits adjustable from 0
to 9999 seconds
Interface for computer
RS 232/V24C ; 300, 600, 1200, 2400, 4800,
9600 bauds (programmable)
TRANSMITTER ENCLOSURE
Material
ABS
Tightness
IP65, according to norm IEC 529
Dimensions
H 144 mm, W 144 mm, D 152 mm
Glands
4 x PG9 - 1 x PG13,5
Material : polyamide
Storage temperature
-20...+70 C
B
Weight
1,7 kg
ENCLOSURE (OPTIONAL)
Standard 19" width for wall or panel
mounting
Material
ABS with PMMA window
Connections
1 gland DN 6/8 for outlet
1 gland DN 4/6 for water input
1 x PG13,5 for mains
1 x PG13,5 for overflow output
6 x PG9 for different connections
Summary of Contents for SODIMAT 9073
Page 1: ...SODIMAT Model 9073 Instruction manual 03 2001 621 190 073 Rev D ...
Page 2: ......
Page 12: ......
Page 20: ......
Page 27: ...SODIMAT Model 9073 3 7 3 3 PREAMPLIFIER Figure 3 4 Preamplifier connections ...
Page 34: ......
Page 36: ...SODIMAT Model 9073 5 2 5 1 General configuration and calibrations ...
Page 38: ...SODIMAT Model 9073 5 4 5 2 Alarm configuration ...
Page 40: ...SODIMAT Model 9073 5 6 5 3 Analogue output configuration ...
Page 42: ...3 2 1 0 0 1 1 0 2 3 4 5 SODIMAT Model 9073 5 8 5 4 Serial interface configuration ...
Page 56: ......
Page 58: ......
Page 69: ...SODIMAT Model 9073 10 8 Figure 10 4 Changing the pump tubing ...
Page 74: ......
Page 80: ......