SODIMAT - Model 9073
3-1
Chapter 3 : Installation
3.1 Unpacking
The analyser is shipped with all its sub-units
factory-adjusted and programmed with
default values. Accessories and options
worth the analyser are factory-mounted
within the appropriate unit.
3.2 Inspection
The SODIMAT 9073 has been inspected
and tested prior to shipment. However, it is
advisable to inspect all parts immediately
upon receipt for any damage which may
have occured during shipment. A damaged
shipping container may indicate internal
damage which may not be immediately
obvious. If there is any evidence of
damage, keep the shipping container.
Any shortage of parts or accessories should
be reported to the authorized Polymetron
distributor or to :
Polymetron S.A.
33, rue du Ballon
93160 Noisy-le-Grand
FRANCE
3.3 Mounting
The instrument requires the following
connections :
- Sample
- Drain
- Power supply
CAUTION !
Mounting should be done by qualified
service personnel only. Any damage
on the instrument by a non qualified
person cancels the warranty
.
3.4 Location
The analyser should be placed in any
convenient location, e.g., installation onto
the sampling rack. The location should be
easily accessible to facilitate the checking
and the regular maintenance.
Before any electrical or hydraulic
connection, check the environment
corresponds to the instrument configuration
:
- voltage
- frequency
- sample supply pressure
- sample temperature
- ambient temperature
Choose a dry site for your instrument if it is
a panel-mounting type instrument. Check
also the quality of the conditioning reagent
: diisopropylamine for the gaseous
conditioning and ethanolamine or ammonia
for the liquid conditioning. These chemical
products should be of “per analysis” quality.
Summary of Contents for SODIMAT 9073
Page 1: ...SODIMAT Model 9073 Instruction manual 03 2001 621 190 073 Rev D ...
Page 2: ......
Page 12: ......
Page 20: ......
Page 27: ...SODIMAT Model 9073 3 7 3 3 PREAMPLIFIER Figure 3 4 Preamplifier connections ...
Page 34: ......
Page 36: ...SODIMAT Model 9073 5 2 5 1 General configuration and calibrations ...
Page 38: ...SODIMAT Model 9073 5 4 5 2 Alarm configuration ...
Page 40: ...SODIMAT Model 9073 5 6 5 3 Analogue output configuration ...
Page 42: ...3 2 1 0 0 1 1 0 2 3 4 5 SODIMAT Model 9073 5 8 5 4 Serial interface configuration ...
Page 56: ......
Page 58: ......
Page 69: ...SODIMAT Model 9073 10 8 Figure 10 4 Changing the pump tubing ...
Page 74: ......
Page 80: ......