SODIMAT - Model 9073
3-2
3.5 Hydraulic connections
3.5.1 CARCAL version (see figure 2-4)
- Fill in the conditioning reagent bottle and
connect it to the circuit.
- Put a filter cartridge on the CARCAL
holder respecting the flow direction shown
on the cartridge.
- Set the sample drain. Caution ! The outlet
is gravitary, it shoud be as direct as
possible (avoid tubing twist).
3.5.2 AUTOCAL version (see figure 2-2)
3
3
MONEC
- Fill in the conditioning reagent bottle and Never attempt to switch on the instrument
connect it to the circuit. before all the connections have been
- Set the sample drain. Caution ! The outlet Before connecting the instrument check the
is gravitary, it shoud be as direct as selected power supply corresponds to your
possible (avoid tubing twist). power supply.
- Fill in the calibration solution bottle with Remove the front panel and the protective
the adequat and fresh solution and put the cover to reach the main selection bridges.
strainer (see table in paragraph 6.2.2 for An aluminuim plate shows the terminal
the concentration used). function and its connection with the external
Caution ! The calibration solution should be
free of solid body which may damage the
calibration pump.
Avoid any direct handling of parts in contact
with the calibration solution : it may cause
pollution.
3.5.3 Mixed bed option
Installing this option requires a qualified
Polymetron technician.
L
See Chapter 9 : Options
3.5.4 Liquid conditioning option (kit K)
L
See Chapter 9 : Options
3.6 Electrical connections
realised.
elements from the monec.
Power supply Min. range Max. range Bridges' Fuses
(V) allowed allowed location mA
- 10 % + 10 %
24 20,4 26,4 X1, X3 T1,25 A
110 94 121,0 X2, X4, X6 T250 mA
127 115 146,0 X1, X4, X6 T250 mA
220 187 242,0 X2, X5 T125 mA
240 207 268,0 X1, X5 T125 mA
Install as many cable glands as necessary. Start with the ground connection.
Summary of Contents for SODIMAT 9073
Page 1: ...SODIMAT Model 9073 Instruction manual 03 2001 621 190 073 Rev D ...
Page 2: ......
Page 12: ......
Page 20: ......
Page 27: ...SODIMAT Model 9073 3 7 3 3 PREAMPLIFIER Figure 3 4 Preamplifier connections ...
Page 34: ......
Page 36: ...SODIMAT Model 9073 5 2 5 1 General configuration and calibrations ...
Page 38: ...SODIMAT Model 9073 5 4 5 2 Alarm configuration ...
Page 40: ...SODIMAT Model 9073 5 6 5 3 Analogue output configuration ...
Page 42: ...3 2 1 0 0 1 1 0 2 3 4 5 SODIMAT Model 9073 5 8 5 4 Serial interface configuration ...
Page 56: ......
Page 58: ......
Page 69: ...SODIMAT Model 9073 10 8 Figure 10 4 Changing the pump tubing ...
Page 74: ......
Page 80: ......