![Polymetron SODIMAT 9073 Instruction Manual Download Page 43](http://html1.mh-extra.com/html/polymetron/sodimat-9073/sodimat-9073_instruction-manual_1578009043.webp)
SODIMAT - Model 9073
5-9
Remark
In order to receive only "one" transmission of the measurement data, select command 820
(1) and print the following command : "R830".
Command
Function
Option
800
0*
No transmission of alarm limits
1
Transmission of S1 and S2 limits
2
Transmission of the alarm system
3
Transmission of the S1 and S2 limits and the alarm system
810
0*
RS232 inactive
1
RS232 active
820
0
Continuous transmission of measuring data
1
Transmission on request only
830*
------------
Transmission of the data measurement (if C820 = 1)
840
0*
300 baud
1
600 baud
2
1200 baud
3
2400 baud
4
4800 baud
5
9600 baud
* : Command accessible only via RS232. It is not displayed..
5.5 Use of the interface RS232
The following codes enable the RS232 interface. Specifications of these codes :
@
Speed in baud (Command 840)
@
1 stop bit
@
No parity bit
@
8 bits per word
The data are in ASCII form :
M1 : 23.00 ppm 25.0 C -360.0 mV
B
The transmitter sends back the data received to allow the reception control.
TRANSFERED CONTROL OF THE SODIMAT
2 characters allow the transfered control : "@" and "!"
the "@" character transfers the control to the computer. The "!" character passes the control to the Monec
keyboard. The control transfer to a remote computer deactivates the transmitter keyboard.
Summary of Contents for SODIMAT 9073
Page 1: ...SODIMAT Model 9073 Instruction manual 03 2001 621 190 073 Rev D ...
Page 2: ......
Page 12: ......
Page 20: ......
Page 27: ...SODIMAT Model 9073 3 7 3 3 PREAMPLIFIER Figure 3 4 Preamplifier connections ...
Page 34: ......
Page 36: ...SODIMAT Model 9073 5 2 5 1 General configuration and calibrations ...
Page 38: ...SODIMAT Model 9073 5 4 5 2 Alarm configuration ...
Page 40: ...SODIMAT Model 9073 5 6 5 3 Analogue output configuration ...
Page 42: ...3 2 1 0 0 1 1 0 2 3 4 5 SODIMAT Model 9073 5 8 5 4 Serial interface configuration ...
Page 56: ......
Page 58: ......
Page 69: ...SODIMAT Model 9073 10 8 Figure 10 4 Changing the pump tubing ...
Page 74: ......
Page 80: ......