SODIMAT - Model 9073
9-2
Measurement not correct
Instrument not programmed properly
Check the programming parameters.
Reprogram the instrument checking it
corresponds to the probe
characteristics
System with probe not calibrated
Calibrate the system with probe
properly
connected.
Probe not connected properly
Check the connection
Probe not operating properly, may be
Check the probe visually. Check the
impossible with this application
probe specifications with the
application
CPU board malfunctionning
If problem persists, call the Service
Technician.
Measurement not stable
Probe malfunctionning
Check the probe
Probe not connected properly
Check the connection
Interferences
Check there in no chemical, external
Cable shield not correctly connected
source
temperature or pressure interference
Check and connect again.
Lack of conditioning reagent
Add reagent
Lack of electrolyte in the reference
Add electrolyte .
electrode
Temperature measurement not
Probe not properly connected
Check it
correct
Temperature not calibrated
Calibrate for temperature
CPU board malfunctionning
If problem persists, call the Service
Technician.
Reading blocked, can not be changed
CPU board and/or other part of
Check the probe is correctly connected
transmitter
Zeroing
Program the instrument again
If it still does not work : reset and
switch off the instrument
Program again
If problem persists, call the Service
Technician.
ERROR displayed
Probe malfunctionning
Check for dirt or electrode
Probe not connected properly
Check the connections
No contact between the probe and
Check the sample flowrate
the sample
CPU board malfunctionning
If problem persists, call the Service
Technician.
Relay not energised
Instrument was programmed
Check whether the correct relay
incorrectly
parameters and setpoints have been
programmed.
Hardware is defective
Check that the programmed setpoints
are compatible with the programmed
measuring range.
Summary of Contents for SODIMAT 9073
Page 1: ...SODIMAT Model 9073 Instruction manual 03 2001 621 190 073 Rev D ...
Page 2: ......
Page 12: ......
Page 20: ......
Page 27: ...SODIMAT Model 9073 3 7 3 3 PREAMPLIFIER Figure 3 4 Preamplifier connections ...
Page 34: ......
Page 36: ...SODIMAT Model 9073 5 2 5 1 General configuration and calibrations ...
Page 38: ...SODIMAT Model 9073 5 4 5 2 Alarm configuration ...
Page 40: ...SODIMAT Model 9073 5 6 5 3 Analogue output configuration ...
Page 42: ...3 2 1 0 0 1 1 0 2 3 4 5 SODIMAT Model 9073 5 8 5 4 Serial interface configuration ...
Page 56: ......
Page 58: ......
Page 69: ...SODIMAT Model 9073 10 8 Figure 10 4 Changing the pump tubing ...
Page 74: ......
Page 80: ......