SODIMAT - Model 9073
6-4
Warning : do not forget to put back the filter after the calibration is
completed.
Note 1 :
- Pressing SELECT interrups the calibration
- Pressing ENTER validates the measurement before the required time (if the measurement
is considered stable).
Note 2 :
- C251 and C261 permit to control the new offset and slope values.
- It is possible to take a note of these values in Appendix B to check the electrode
evolution.
The analyser is then able to calculate the new offset and slope.
If any big difference after the calculation compared to the theoric values, the following error
message is displayed.
6.3.2 AUTOCAL calibration
The sodium concentration of the calibration solution to be prepared depends of the sodium
contained in the water to analyse. The values advised below will ensure a utmost accuracy :
Water to analyse
Calibration
Addition 1 generated
Addition 2 generated
(ppb Na*)
solution
(ppb) flow rate = 4 l/h
(ppb) flow rate = 4 l/h
< 0.5 ppb
10 ppm
954
87.1
< 1.9 ppb
1 ppm
95.4
8.71
< 3.8 ppb
2 ppm
190.8
17.42
< 9 ppb
5 ppm
477
43.56
< 38 ppb
20 ppm
1908
174.2
< 90 ppb
50 ppm
4770
435.6
(
Preparation of a 2 ppm Na calibration solution :
ì
Prepare a stock solution 1 g/l Na (1000 ppm) :
Weigh 2.541 g of NaCl dry and dissolve it in one litre demineralised water, free of sodium.
This solution should be kept in a cool and dry place, its lifetime is 6 months.
í
From the stock solution :
Take 2 ml of the stock solution and dilute it in one litre demineralised water to obtain a 2
ppm Na calibration solution.
Summary of Contents for SODIMAT 9073
Page 1: ...SODIMAT Model 9073 Instruction manual 03 2001 621 190 073 Rev D ...
Page 2: ......
Page 12: ......
Page 20: ......
Page 27: ...SODIMAT Model 9073 3 7 3 3 PREAMPLIFIER Figure 3 4 Preamplifier connections ...
Page 34: ......
Page 36: ...SODIMAT Model 9073 5 2 5 1 General configuration and calibrations ...
Page 38: ...SODIMAT Model 9073 5 4 5 2 Alarm configuration ...
Page 40: ...SODIMAT Model 9073 5 6 5 3 Analogue output configuration ...
Page 42: ...3 2 1 0 0 1 1 0 2 3 4 5 SODIMAT Model 9073 5 8 5 4 Serial interface configuration ...
Page 56: ......
Page 58: ......
Page 69: ...SODIMAT Model 9073 10 8 Figure 10 4 Changing the pump tubing ...
Page 74: ......
Page 80: ......